用于天然气等工艺气体中微量水分(露点)的在线检测。
基本参数
| 技术参数 | |||
| 测量气体① | 量程 | 精度② | 分辨率 |
| O2 | 0~10000ppm | ±2%F.S. | 1ppm |
| 0~5%VOL | ±2%F.S. | 0.01% | |
| 0~25%VOL | ±1%F.S. | ||
| 0~100%VOL | ±0.5%F.S. | ||
| CO | 0~500ppm | ±2%F.S. | 0.1ppm |
| 0~1000ppm | ±1%F.S. | ||
| 0~5%VOL | ±2%F.S. | 0.01% | |
| 0~25%VOL | ±1%F.S. | ||
| 0~100%VOL | ±1%F.S. | ||
| CO2 | 0~1000ppm | ±2%F.S. | 0.1ppm |
| 0~5%VOL | ±2%F.S. | 0.01% | |
| 0~25%VOL | ±1%F.S. | ||
| 0~100%VOL | ±1%F.S. | ||
| H2S | 0~1000ppm | ±2%F.S. | 0.1ppm |
| 0~1%VOL | ±2%F.S. | 0.01% | |
| CH4 | 0~100ppm | ±2%F.S. | 0.1ppm |
| 0~1000ppm | ±1%F.S. | ||
| 0~5%VOL | ±2%F.S. | 0.01% | |
| 0~25%VOL | ±1%F.S. | ||
| 0~100%VOL | ±1%F.S. | ||
| 测量原理 | 可调谐半导体激光吸收光谱技术(TDLAS ) | ||
| 响应时间③ | <40s | ||
| 最佳流量④ | 0.7~1.2L/min | ||
| 样气要求 | 无尘、无水、无油 | ||
| 电气参数 | |||
| 通信 | RS-485/RS-232;4~20mA | ||
| 电源 | AC220V,50Hz | ||
| 显示 | LCD显示 | ||
| 防爆等级 | Ex db IIC T6 Gb / Ex tb IIIC T80℃ Db | ||
| 功能配置 | 内置调零气泵,可以实现空气自动调零;具备自诊断功能,可在线检查传感器状态 | ||
| ① 表格所有数据是在标准??压下测量。 ② 测量标准?体由国家计量部?认证的单位提供。在?燥氮?中稀释的所有?体;氮?要求精度?于99.999%。 ③ 响应时间与进?流量相关,该响应时间为2L/min流量下测试。 ④ 最佳流量为0.7~1.2L/min,实际应用时可达2L/min | |||
应用领域
适用于天然气开采/运输/存储/液化加工、精制炼油厂、石油石化、垃圾填埋等诸多领域的工业气体排放监测和过程控制。